IEC 62353:2014 (2014-09) |
MEDICAL ELECTRICAL EQUIPMENT - RECURRENT TEST AND TEST AFTER REPAIR OF MEDICAL ELECTRICAL EQUIPMENT |
|
Price Excl. VAT |
Total number of pages, tables and drawings |
366.00 €
|
129 P. |
Description
IEC 62353:2014 applies to testing of medical electrical equipment and medical electrical systems, hereafter referred to as ME equipment and ME systems, or parts of such equipment or systems, which comply with IEC 60601-1:1988 (second edition) and its amendments and IEC 60601-1:2005 (third edition) and its amendments, before putting into service, during maintenance, inspection, servicing and after repair or on occasion of recurrent tests to assess the safety of such ME equipment or ME systems or parts thereof. For equipment not built to IEC 60601-1 these requirements may be used taking into account the safety standards for the design and information in the instructions for use of that equipment. This standard contains tables with allowable values relating to different editions of IEC 60601-1. For the purpose of this standard, the application of measuring methods is independent of the edition according to which the ME equipment or ME system is designed. This standard contains "general requirements", which contain clauses of general concern, and "particular requirements", further clauses handling special types of ME equipment or ME systems and applying in connection with the "General requirements". This standard is not suitable to assess whether ME equipment or ME systems or any other equipment comply with the relevant standards for their design. This standard is not applicable to the assembly of ME systems. For assembling ME systems see Clause 16 of IEC 60601-1:2005 + IEC 60601-1:2005/AMD1:2012. This standard does not define requirements for repair, exchange of components and modification of ME equipment or ME systems. All maintenance, inspection, servicing, and repair done in accordance with the manufacturer's instructions maintain the conformity to the standard used for the design of the equipment. Otherwise conformity to applicable requirements should be assessed and verified, before the tests of this standard are performed. This standard is also applicable to tests after repair. This second edition cancels and replaces the first edition of IEC 62353 published in 2007. This edition constitutes a technical revision including the following main revisions:
- clarification in 5.3.4.1 that measurements of leakage currents based on test configurations derived from IEC 60601-1 are an allowable alternative method and the inclusion of informative explanation in Annex A;
- revision of the protective earth resistance requirements for ME systems using multiple socket outlets to take account of IEC 60601-1:2005/AMD1:2012 on the safe allowed values of protective earth resistance of plugged-in equipment;
- the inclusion of expected minimum insulation resistance values in Table 2; and
- a reordering of the sequence of testing in Annex B.
|
Class |
C990
(IEC PUBLICATIONS IEC PUBLICATIONS)
|
Status |
IEC PUBLICATION |
Situation |
Currently active
Replaces
IEC 62353:2007
|
|
Committee |
TC 62/SC 62A
COMMON ASPECTS OF ELECTRICAL EQUIPMENT USED IN MEDICAL PRACTICE
|
Responsible |
De heer VAN HECKE Luk
|
BEC Approval |
2014-09-04 |
ICS-Code (International Standards Classification) |
11.040.01
|
NBN Status |
New |
|
IEC publication date |
2014-09-04 |
IEC stability date |
2022-12-31 |
IEC file modification date |
2014-09-03 |
IEC last modification date |